A new polarographic system for measurement of oxygen consumption in vitro
has been constructed to minimize different types of errors. The system was
built of polyamide-6, Monax glass, stainless steel, and rubber to diminish
oxygen leakage and gas capacitance properties. With these materials the
system was gas-tight. Nevertheless, the incubation chambers had oxygen
capacitance properties that had to be corrected for mathematically. Thus,
in order to obtain more exact absolute values of the respiratory rate, the
apparent respiratory slopes were continuously corrected for delta pO2.
Furthermore, the initial slope observed in the absence of biological
material was also taken into account. Using these variables, the apparent
slope (mm Hg/min) could be adjusted by subtracting the following correction
factor: correction factor = 0.00306 X delta pO2 + 0.198 X (initial slope).
This correction gave a high degree of linearity and an overall error on the
order of 10%. Without the described correction for the oxygen capacitance,
the true respiratory rate may be underestimated by up to 50%.
ARTICLES
Measurements of adipose tissue respiration in a closed chamber using an oxygen sensor: methodological considerations
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
J. F. Himm and L. D. Homer A model of extravascular bubble evolution: effect of changes in breathing gas composition J Appl Physiol, October 1, 1999; 87(4): 1521 - 1531. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| All ASBMB Journals | Journal of Biological Chemistry |
| Molecular and Cellular Proteomics | ASBMB Today |