Journal of Lipid Research, Vol. 48, 458-464, February 2007
Copyright © 2007 by American Society for Biochemistry and Molecular Biology
Methods |
-hydroxy-4-cholesten-3-one in human serum by LC-ESI-MS/MS




* Ibaraki Prefectural Institute of Public Health, Mito, Ibaraki 310-0852, Japan
Faculty of Pharmaceutical Science, Tohoku Pharmaceutical University, Sendai, Miyagi 981-8558, Japan
Department of Medicine, Tokyo Medical University, Kasumigaura Hospital, Ami, Ibaraki 300-0395, Japan
** Pharmax Institute, Kawasaki, Kanagawa 213-0021, Japan
Published, JLR Papers in Press, November 8, 2006.
1 To whom correspondence should be addressed. e-mail: akihonda-gi{at}umin.ac.jp
| ABSTRACT |
|---|
|
|
|---|
-hydroxy-4-cholesten-3-one (C4), which has been used as a biomarker for bile acid biosynthesis. This method is based upon a stable isotope dilution technique by liquid chromatography-tandem mass spectrometry (LC-MS/MS). C4 was extracted from human serum (250 µl) by a salting-out procedure, derivatized into the picolinoyl ester (C4-7
-picolinate), and then purified using a disposable C18 cartridge. The resulting picolinoyl ester derivative of C4 was quantified by LC-MS/MS using the electrospray ionization mode. The detection limit of the C4 picolinoyl ester was found to be 100 fg (signal-to-noise ratio = 10), which was
1,000 times more sensitive than the detection limit of C4 with a conventional HPLC-ultraviolet method. The relative standard deviations between sample preparations and between measurements by our method were calculated to be 5.7% and 3.9%, respectively, by one-way layout analysis. The recovery experiments were performed using serum spiked with 20.060.0 ng/ml C4 and were validated by a polynomial equation. The results showed that the estimated concentration with 95% confidence limit was 23.1 ± 2.8 ng/ml, which coincided completely with the observed
0 ± SD = 23.3 ± 1.0 ng/ml with a mean recovery of 93.4%. This method provides highly reliable and reproducible results for the quantification of C4, especially in small volumes of blood samples.
Supplementary key words bile acid synthesis cholesterol 7
-hydroxylase picolinic acid liquid chromatography electrospray ionization-tandem mass spectrometry
Abbreviations: C4, 7
-hydroxy-4-cholesten-3-one; CYP7A1, cholesterol 7
-hydroxylase; ESI, electrospray ionization; GC-SIM, gas chromatography-mass spectrometry with selected ion monitoring; LC-MS/MS, liquid chromatography-tandem mass spectrometry; SRM, selected reaction monitoring; UV, ultraviolet
| INTRODUCTION |
|---|
|
|
|---|
-hydroxylase (CYP7A1) is the rate-limiting enzyme in one of the pathways of bile acid biosynthesis (2). Bile acid synthesis by this classic pathway accounts for <50% of total bile acid production in rats (3) and mice (4), whereas it represents a major route synthesizing >90% of total bile acids in humans (5). Therefore, the measurement of CYP7A1 activity is clinically useful for exploring the mechanisms of hypercholesterolemia (6) and evaluating the effects of hypocholesterolemic treatments (7, 8).
Because CYP7A1 is expressed solely in the liver, an invasive liver biopsy is necessary to measure the activity of this enzyme. However, a breakthrough of this problem was achieved in 1987 by Björkhem et al. (9). They quantified the plasma concentrations of 7
-hydroxycholesterol, the immediate product of CYP7A1, by gas chromatography-mass spectrometry with selected ion monitoring (GC-SIM) and reported that this measurement reflected hepatic CYP7A1 activities in humans.
In 1988, Axelson, Aly, and Sjövall (10) determined the plasma concentrations of 7
-hydroxy-4-cholesten-3-one (C4), a product of the next oxidative enzymatic reaction after CYP7A1, using HPLC with ultraviolet (UV) detection and found out that C4 is also a good marker for CYP7A1 activity in humans (11). It was subsequently reported that serum concentrations of 7
-hydroxycholesterol (12) and C4 (13) reflected not only CYP7A1 activities but also bile acid synthesis in humans.
We also developed new high-resolution GC-SIM methods for the quantification of these sterols in plasma. We found that there was a significant correlation between hepatic CYP7A1 activities and plasma levels of free but not esterified or total 7
-hydroxycholesterol (14) and that CYP7A1 activities correlated better with plasma levels of C4 than those of free 7
-hydroxycholesterol (15). In addition, we applied these methods to monitor the circadian rhythms of these sterol levels in human plasma. A zero-amplitude test (15) revealed that C4 was the most reliable plasma biomarker for hepatic CYP7A1 activity. Furthermore, our recent study demonstrated that plasma C4 relative to cholesterol is a better marker for hepatic CYP7A1 activity than the absolute concentration when hypercholesterolemia is present (16).
The aim of this study was to develop a simple, more sensitive and reliable method to quantify C4 in a small volume of human serum. For this purpose, we derivatized C4 into the picolinoyl ester (C4-7
-picolinate) and determined the concentrations using liquid chromatography-tandem mass spectrometry (LC-MS/MS) coupled in electrospray ionization (ESI) mode.
| MATERIALS AND METHODS |
|---|
|
|
|---|
-hydroxycholesterol was prepared as described previously (17). [2H7]C4 and 7ß-hydroxy-4-cholesten-3-one were synthesized from [2H7]7
-hydroxycholesterol and 7ß-hydroxycholesterol, respectively, with cholesterol oxidase (18). Picolinic acid and 2-methyl-6-nitrobenzoic anhydride were purchased from Tokyo Kasei Kogyo (Tokyo, Japan), and 4-dimethylaminopyridine and triethylamine were obtained from Wako Pure Chemical Industries (Osaka, Japan). Additional reagents and solvents were of analytical grade.
Sample collection
Blood samples were collected from healthy human volunteers. After coagulation and centrifugation at 1,500 g for 10 min, serum samples were stored at 20°C until analysis. Informed consent was obtained from all subjects, and the experimental procedures were conducted in accordance with the ethical standards of the Helsinki Declaration.
Sample preparation
C4 was extracted from serum by the salting-out method described previously (15). Fifty microliters of serum was diluted with 200 µl of distilled water in a microcentrifuge tube (1.5 ml; Eppendorf, Hamburg, Germany), and 1 ng of [2H7]C4 in 500 µl of acetonitrile was added as an internal standard. After the addition of
100 mg of ammonium sulfate, the sample tube was vortexed for 1 min and centrifuged at 2,000 g for 5 min. The supernatant acetonitrile phase was collected and evaporated to dryness under nitrogen.
Derivatization to the picolinoyl ester was performed according to the method of Yamashita et al. (19) with minor modifications. The reagent mixture for derivatization consisted of 2-methyl-6-nitrobenzoic anhydride (100 mg), 4-dimethylaminopyridine (30 mg), picolinic acid (80 mg), and tetrahydrofuran (1.5 ml). The freshly prepared reagent mixture (150 µl) and triethylamine (20 µl) were added to the extract, and the reaction mixture was allowed to stand at room temperature for 30 min. After the addition of 0.5 ml of 5% sodium bicarbonate, the mixture was applied to a Bond Elut C18 cartridge (100 mg; Varian, Harbor City, CA) preconditioned with 1 ml of tetrahydrofuran, 1 ml of methanol, and 2 ml of distilled water. The cartridge was washed with 1 ml of 5% sodium bicarbonate, 1 ml of distilled water, 2 ml of 5% HCl, 1 ml of distilled water again, and 2 ml of acetonitrile-water (1:1, v/v). The picolinoyl ester derivative of C4 was then eluted with 4 ml of acetonitrile-water (95:5, v/v) containing 0.1% acetic acid. After evaporation, the residue was redissolved in 50 µl of the same solvent, and an aliquot (1 µl) was injected into the LC-MS/MS system.
LC-MS/MS analysis
The LC-MS/MS system consisted of a TSQ Quantum Ultra quadrupole mass spectrometer (Thermo Fisher Scientific, San Jose, CA) equipped with an H-ESI probe and a Nanospace SI-2 HPLC system (Shiseido, Tokyo, Japan). Chromatographic separation was performed using a Hypersil GOLD column (150 x 2.1 mm, 3 µm; Thermo Fisher Scientific) at 40°C. The mobile phase consisted of acetonitrile-water (95:5, v/v) containing 0.1% acetic acid and was used at a flow rate of 200 µl/min. The general LC-MS/MS conditions were as follows: spray voltage, 1,000 V; vaporizer temperature, 350°C; sheath gas (nitrogen) pressure, 35 p.s.i.; auxiliary gas (nitrogen) flow, 45 arbitrary units; ion transfer capillary temperature, 350°C; collision gas (argon) pressure, 1.5 mTorr; collision energy, 16 V; and ion polarity, positive.
Evaluation of the solid-phase purification process
The absolute recovery of C4-7
-picolinate from the Bond Elut C18 cartridge was determined by the addition of [2H7]C4-7
-picolinate (1 ng) to the cartridge with a derivatizing mixture, which was prepared from 50 µl of serum without adding [2H7]C4. After the purification steps, the picolinoyl ester of 7ß-hydroxy-4-cholesten-3-one (1 ng) was added to the eluate as an internal standard, and both [2H7]C4-7
-picolinate and 7ß-hydroxy-4-cholesten-3-one-7ß-picolinate were quantified by LC-selected reaction monitoring (SRM).
To calculate the elimination rate of cholesterol by the Bond Elut C18 cartridge, unesterified cholesterol concentration in the first salting-out extraction was measured by colorimetric enzyme assay. The extract was then derivatized and purified by the cartridge, and the picolinoyl ester of [2H7]cholesterol (100 ng) was added to the final eluate as an internal standard. Cholesterol-3-picolinate and its 2H7 variant were quantified by the LC-SIM method to measure the absolute amount of cholesterol-3-picolinate contaminant in the final eluate.
Statistics
Data are reported as means ± SD. Reproducibility was analyzed by one-way layout (JMP software; SAS Institute, Inc., Cary, NC). Recovery was analyzed using a polynomial equation (20). Linearity of the calibration curve and correlation between serum volume and the amount of C4 were analyzed by simple linear regression. The regression analysis was also used to calculate the estimated amount ± 95% confidence limit in the recovery study. For all analyses, significance was accepted at P < 0.05.
| RESULTS |
|---|
|
|
|---|
-picolinate is shown in Fig. 1A
. This picolinoyl ester derivative exhibited an [M+H]+ ion at m/z 506 as the base peak. In the MS/MS spectrum, the [MC6H5O2N]+ ion was observed at m/z 383 as the most prominent peak, as shown in Fig. 1B. The SRM was conducted using m/z 506
m/z 383 for the C4-7
-picolinate and m/z 513
m/z 390 for the 2H7 variant.
|
-picolinate solution was diluted and injected into the LC-MS/MS system. As shown in Fig. 2A
, the C4-7
-picolinate was easily detected to 100 fg, with a signal-to-noise ratio of 10, whereas conventional UV detection at 241 nm was barely able to detect 100 pg of C4 (Fig. 2B).
|
-picolinate was used instead of the corresponding C4 derivative and was quantified by SRM using the picolinoyl ester of 7ß-hydroxy-4-cholesten-3-one as an internal standard. Both picolinoyl esters of C4 and 7ß-hydroxy-4-cholesten-3-one exhibited similar mass spectra, but the retention times were 4.66 and 5.00 min, respectively. Thus, the recovery of [2H7]C4-7
-picolinate was 104.0 ± 2.3% (n = 4).
The elimination rate of cholesterol by the Bond Elut C18 cartridge was calculated by quantifying cholesterol-3-picolinate contamination in the final eluate. The multiple-ion detector was focused on m/z 492 for cholesterol-3-picolinate and m/z 499 for [2H7]cholesterol-3-picolinate. Under the same HPLC conditions as for the determination of C4, the retention time of cholesterol-3-picolinate was
11 min. The elimination rate of cholesterol by the cartridge was calculated as 98.6 ± 0.7% (n = 4).
Calibration curve
A calibration curve was established for C4 (Fig. 3
). Different amounts (0.01, 0.05, 0.1, 0.25, 0.5, 1, 2, and 4 ng) of authentic C4 were mixed with 1 ng of [2H7]C4, derivatized to the picolinoyl ester, and purified by the Bond Elut C18 cartridge, as described in Materials and Methods. The weight ratio of C4, relative to the corresponding deuterated internal standard, was plotted on the abscissa, and the peak area ratio of the C4-7
-picolinate to the 2H7 variant, as measured by LC-SRM, was plotted on the ordinate. The linearity of the standard curve, as determined by simple linear regression, was excellent for weight ratios between 0.01 and 4.0 (n = 8; r = 1.000; P < 0.0001).
|
-picolinate and the 2H7 variant obtained from 50 µl of normal human serum. Both peaks corresponded to
20 pg. The peak appearing at 5.46 min on this chromatogram (Fig. 4A) was identified as the picolinoyl ester derivative of 7-oxo-cholesterol by comparison with authentic 7-oxo-cholesterol-3-picolinate.
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
We previously developed a more sensitive and specific method based on a stable isotope dilution technique using high-resolution GC-SIM (15). In this method, C4 was extracted from 200 µl of human plasma by a salting-out extraction, purified by a commercially available solid-phase extraction column (C18 and unbonded silica) at room temperature, and converted to the methyloxime-dimethylethylsilyl ether derivative before GC-SIM analysis. However, the sensitivity of this method was still not sufficient for the quantification of C4 in limited amounts of human serum, because the resulting derivatives exhibited two peaks of syn and anti isomers and their fragmentation patterns were not simple.
LC-MS or LC-MS/MS has come to be used more readily than GC-MS, because these methods do not always require a derivatization step. However, it is also true that the introduction of charged moieties enhances the ionization efficiency of neutral steroids in ESI and atmospheric pressure chemical ionization processes. A permanently charged N-methylpyridyl group was introduced into the hydroxyl group of cholesterol (23) and 5
-dihydrotestosterone (24, 25), whereas a permanently charged quaternary pyridinium moiety was introduced to the carbonyl group of oxosteroids with Girard P reagent (2628) or 2-hydrazino-1-methylpyridine (29). Very recently, Yamashita et al. (19) developed a new picolinoyl derivatization that is a simple and versatile method suitable for the sensitive and specific quantification of hydroxysteroids by positive ESI-LC-MS/MS. Interestingly, because the picolinoyl group is not charged permanently, even in the case of estradiol with two hydroxyl groups, a single charged ion was observed in the positive ESI mass spectrum. Therefore, picolinoyl ester may be a suitable derivative not only for monohydroxy steroids but also for hydroxycholesterols with plural hydroxyl groups.
Then, we applied this derivatization to quantify C4 in a very small volume of human serum by LC-SRM. The picolinoyl ester of C4 exhibited [M+H]+ as a single ion in ESI mode. The detection limit of C4 (as the C4 picolinoyl ester) in SRM mode was 100 fg, which was
1,000 and 30 times more sensitive than that by the HPLC-UV and high-resolution GC-SIM methods, respectively. The detection limit of 100 fg corresponds to 0.1 ng/ml in 50 µl of serum, whereas the range of serum C4 concentrations has been reported to be 340 ng/ml (median, 12 ng/ml) in healthy subjects (10). Consequently, even when a sufficient volume of serum was not available, C4 concentrations could be determined by our method using a few microliters of serum. However, we recommend that 50 µl of serum be used for routine assay because some subjects treated with chenodeoxycholic acid (11) and patients with CYP7A1 deficiency (6) appear to exhibit unusually low serum concentrations of C4.
The derivatization and purification steps in this method are very simple. After the previously described salting-out extraction procedure (15), hydroxysteroids, including C4 and cholesterol, were derivatized to the picolinoyl esters within 30 min at room temperature. The C4-7
-picolinate was completely recovered from a commercially available C18 column with >98% elimination of cholesterol-3-picolinate. However, the amount of cholesterol-3-picolinate contaminant in the final eluate was still 300 times greater than that of C4-7
-picolinate. Therefore, the next injection into the LC-MS/MS system was not performed until cholesterol-3-picolinate had eluted completely from the column (
12 min), whereas the retention time of C4-7
-picolinate was
4.7 min.
[2H7]C4 was added to serum as an ideal internal standard for quantification by SRM. Although the object of our study was to quantify a very small amount of C4, the specificity and reproducibility of this method were highly satisfactory.
In summary, a very sensitive and specific method for the quantification of C4 in human serum was developed. Derivatization of C4 into the picolinoyl ester made it possible to be quantified by LC-ESI-MS/MS with excellent sensitivity. Recovery and reproducibility experiments verified that this method provided analytical results with high reliability and reproducibility.
| ACKNOWLEDGMENTS |
|---|
Manuscript received August 28, 2006 and in revised form October 31, 2006.
| REFERENCES |
|---|
|
|
|---|
-hydroxylase gene in mice. J. Biol. Chem. 271: 1802418031.
-hydroxylase (CYP7A1) deficiency has a hypercholesterolemic phenotype. J. Clin. Invest. 110: 109117.[CrossRef][Medline]
-hydroxycholesterol as a marker for increased activity of the cholesterol 7
-hydroxylase in humans. J. Lipid Res. 28: 889894.[Abstract]
-hydroxy-4-cholesten-3-one in plasma reflect rates of bile acid synthesis in man. FEBS Lett. 239: 324328.[CrossRef][Medline]
-hydroxy-4-cholesten-3-one reflects the activity of hepatic cholesterol 7
-hydroxylase in man. FEBS Lett. 284: 216218.[CrossRef][Medline]
-hydroxycholesterol as an indicator of bile acid synthesis in humans. J. Lipid Res. 36: 20592066.[Abstract]
-hydroxy-4-cholesten-3-one reflect bile acid synthesis in humans. Hepatology. 24: 123126.
-hydroxycholesterol in human plasma by gas chromatography-mass spectrometry as indices of cholesterol and bile acid biosynthesis. J. Chromatogr. 613: 185193.[Medline]
-hydroxy-4-cholesten-3-one level in plasma using isotope-dilution mass spectrometry and monitoring its circadian rhythm in human as an index of bile acid biosynthesis. J. Chromatogr. 655: 179187.
-hydroxy-4-cholesten-3-one and 27-hydroxycholesterol concentrations as markers for hepatic bile acid synthesis in cholesterol-fed rabbits. Metabolism. 53: 4248.[CrossRef][Medline]
-diol and the enzymatic preparation of the labeled cholest-4-en-7
-ol-3-one. Hiroshima J. Med. Sci. 23: 265272.[Medline]
-hydroxy-4-cholesten-3-one in human serum. J. Chromatogr. 657: 3136.[Medline]
-hydroxylase activity by assay of the stable bile acid intermediate 7
-hydroxy-4-cholesten-3-one in peripheral blood. J. Lipid Res. 44: 859865.
-dihydrotestosterone and sensitive analysis by semimicro-LC/ESI-MS. J. Mass Spectrom. Soc. Jpn. 50: 330336.This article has been cited by other articles:
![]() |
A. Honda, K. Yamashita, H. Miyazaki, M. Shirai, T. Ikegami, G. Xu, M. Numazawa, T. Hara, and Y. Matsuzaki Highly sensitive analysis of sterol profiles in human serum by LC-ESI-MS/MS J. Lipid Res., September 1, 2008; 49(9): 2063 - 2073. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Lenicek, M. Juklova, J. Zelenka, J. Kovar, M. Lukas, M. Bortlik, and L. Vitek Improved HPLC Analysis of Serum 7{alpha}-Hydroxycholest-4-en-3-one, a Marker of Bile Acid Malabsorption Clin. Chem., June 1, 2008; 54(6): 1087 - 1088. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||