Originally published In Press as doi:10.1194/jlr.D600032-JLR200 on November 8, 2006

This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
D600032-JLR200v1
48/2/458    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Honda, A.
Right arrow Articles by Miyazaki, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Honda, A.
Right arrow Articles by Miyazaki, H.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

Journal of Lipid Research, Vol. 48, 458-464, February 2007
Copyright © 2007 by American Society for Biochemistry and Molecular Biology


Methods

Highly sensitive quantification of 7{alpha}-hydroxy-4-cholesten-3-one in human serum by LC-ESI-MS/MS

Akira Honda1,*, Kouwa Yamashita{dagger}, Mitsuteru Numazawa{dagger}, Tadashi Ikegami§, Mikio Doy*, Yasushi Matsuzaki§ and Hiroshi Miyazaki**

* Ibaraki Prefectural Institute of Public Health, Mito, Ibaraki 310-0852, Japan
{dagger} Faculty of Pharmaceutical Science, Tohoku Pharmaceutical University, Sendai, Miyagi 981-8558, Japan
§ Department of Medicine, Tokyo Medical University, Kasumigaura Hospital, Ami, Ibaraki 300-0395, Japan
** Pharmax Institute, Kawasaki, Kanagawa 213-0021, Japan

Published, JLR Papers in Press, November 8, 2006.

1 To whom correspondence should be addressed. e-mail: akihonda-gi{at}umin.ac.jp


    ABSTRACT
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
We describe a highly sensitive and specific method for the quantification of serum 7{alpha}-hydroxy-4-cholesten-3-one (C4), which has been used as a biomarker for bile acid biosynthesis. This method is based upon a stable isotope dilution technique by liquid chromatography-tandem mass spectrometry (LC-MS/MS). C4 was extracted from human serum (2–50 µl) by a salting-out procedure, derivatized into the picolinoyl ester (C4-7{alpha}-picolinate), and then purified using a disposable C18 cartridge. The resulting picolinoyl ester derivative of C4 was quantified by LC-MS/MS using the electrospray ionization mode. The detection limit of the C4 picolinoyl ester was found to be 100 fg (signal-to-noise ratio = 10), which was ~1,000 times more sensitive than the detection limit of C4 with a conventional HPLC-ultraviolet method. The relative standard deviations between sample preparations and between measurements by our method were calculated to be 5.7% and 3.9%, respectively, by one-way layout analysis. The recovery experiments were performed using serum spiked with 20.0–60.0 ng/ml C4 and were validated by a polynomial equation. The results showed that the estimated concentration with 95% confidence limit was 23.1 ± 2.8 ng/ml, which coincided completely with the observed X0 ± SD = 23.3 ± 1.0 ng/ml with a mean recovery of 93.4%. This method provides highly reliable and reproducible results for the quantification of C4, especially in small volumes of blood samples.

Supplementary key words bile acid synthesis • cholesterol 7{alpha}-hydroxylase • picolinic acid • liquid chromatography electrospray ionization-tandem mass spectrometry

Abbreviations: C4, 7{alpha}-hydroxy-4-cholesten-3-one; CYP7A1, cholesterol 7{alpha}-hydroxylase; ESI, electrospray ionization; GC-SIM, gas chromatography-mass spectrometry with selected ion monitoring; LC-MS/MS, liquid chromatography-tandem mass spectrometry; SRM, selected reaction monitoring; UV, ultraviolet


    INTRODUCTION
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
The conversion of cholesterol to bile acids is a major reaction for the catabolism of cholesterol in the body (1). Cholesterol 7{alpha}-hydroxylase (CYP7A1) is the rate-limiting enzyme in one of the pathways of bile acid biosynthesis (2). Bile acid synthesis by this classic pathway accounts for <50% of total bile acid production in rats (3) and mice (4), whereas it represents a major route synthesizing >90% of total bile acids in humans (5). Therefore, the measurement of CYP7A1 activity is clinically useful for exploring the mechanisms of hypercholesterolemia (6) and evaluating the effects of hypocholesterolemic treatments (7, 8).

Because CYP7A1 is expressed solely in the liver, an invasive liver biopsy is necessary to measure the activity of this enzyme. However, a breakthrough of this problem was achieved in 1987 by Björkhem et al. (9). They quantified the plasma concentrations of 7{alpha}-hydroxycholesterol, the immediate product of CYP7A1, by gas chromatography-mass spectrometry with selected ion monitoring (GC-SIM) and reported that this measurement reflected hepatic CYP7A1 activities in humans.

In 1988, Axelson, Aly, and Sjövall (10) determined the plasma concentrations of 7{alpha}-hydroxy-4-cholesten-3-one (C4), a product of the next oxidative enzymatic reaction after CYP7A1, using HPLC with ultraviolet (UV) detection and found out that C4 is also a good marker for CYP7A1 activity in humans (11). It was subsequently reported that serum concentrations of 7{alpha}-hydroxycholesterol (12) and C4 (13) reflected not only CYP7A1 activities but also bile acid synthesis in humans.

We also developed new high-resolution GC-SIM methods for the quantification of these sterols in plasma. We found that there was a significant correlation between hepatic CYP7A1 activities and plasma levels of free but not esterified or total 7{alpha}-hydroxycholesterol (14) and that CYP7A1 activities correlated better with plasma levels of C4 than those of free 7{alpha}-hydroxycholesterol (15). In addition, we applied these methods to monitor the circadian rhythms of these sterol levels in human plasma. A zero-amplitude test (15) revealed that C4 was the most reliable plasma biomarker for hepatic CYP7A1 activity. Furthermore, our recent study demonstrated that plasma C4 relative to cholesterol is a better marker for hepatic CYP7A1 activity than the absolute concentration when hypercholesterolemia is present (16).

The aim of this study was to develop a simple, more sensitive and reliable method to quantify C4 in a small volume of human serum. For this purpose, we derivatized C4 into the picolinoyl ester (C4-7{alpha}-picolinate) and determined the concentrations using liquid chromatography-tandem mass spectrometry (LC-MS/MS) coupled in electrospray ionization (ESI) mode.


    MATERIALS AND METHODS
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
Chemicals
C4 and 7ß-hydroxycholesterol were purchased from Steraloids (Wilton, NH). [25,26,26,26,27,27,27-2H7]7{alpha}-hydroxycholesterol was prepared as described previously (17). [2H7]C4 and 7ß-hydroxy-4-cholesten-3-one were synthesized from [2H7]7{alpha}-hydroxycholesterol and 7ß-hydroxycholesterol, respectively, with cholesterol oxidase (18). Picolinic acid and 2-methyl-6-nitrobenzoic anhydride were purchased from Tokyo Kasei Kogyo (Tokyo, Japan), and 4-dimethylaminopyridine and triethylamine were obtained from Wako Pure Chemical Industries (Osaka, Japan). Additional reagents and solvents were of analytical grade.

Sample collection
Blood samples were collected from healthy human volunteers. After coagulation and centrifugation at 1,500 g for 10 min, serum samples were stored at –20°C until analysis. Informed consent was obtained from all subjects, and the experimental procedures were conducted in accordance with the ethical standards of the Helsinki Declaration.

Sample preparation
C4 was extracted from serum by the salting-out method described previously (15). Fifty microliters of serum was diluted with 200 µl of distilled water in a microcentrifuge tube (1.5 ml; Eppendorf, Hamburg, Germany), and 1 ng of [2H7]C4 in 500 µl of acetonitrile was added as an internal standard. After the addition of ~100 mg of ammonium sulfate, the sample tube was vortexed for 1 min and centrifuged at 2,000 g for 5 min. The supernatant acetonitrile phase was collected and evaporated to dryness under nitrogen.

Derivatization to the picolinoyl ester was performed according to the method of Yamashita et al. (19) with minor modifications. The reagent mixture for derivatization consisted of 2-methyl-6-nitrobenzoic anhydride (100 mg), 4-dimethylaminopyridine (30 mg), picolinic acid (80 mg), and tetrahydrofuran (1.5 ml). The freshly prepared reagent mixture (150 µl) and triethylamine (20 µl) were added to the extract, and the reaction mixture was allowed to stand at room temperature for 30 min. After the addition of 0.5 ml of 5% sodium bicarbonate, the mixture was applied to a Bond Elut C18 cartridge (100 mg; Varian, Harbor City, CA) preconditioned with 1 ml of tetrahydrofuran, 1 ml of methanol, and 2 ml of distilled water. The cartridge was washed with 1 ml of 5% sodium bicarbonate, 1 ml of distilled water, 2 ml of 5% HCl, 1 ml of distilled water again, and 2 ml of acetonitrile-water (1:1, v/v). The picolinoyl ester derivative of C4 was then eluted with 4 ml of acetonitrile-water (95:5, v/v) containing 0.1% acetic acid. After evaporation, the residue was redissolved in 50 µl of the same solvent, and an aliquot (1 µl) was injected into the LC-MS/MS system.

LC-MS/MS analysis
The LC-MS/MS system consisted of a TSQ Quantum Ultra quadrupole mass spectrometer (Thermo Fisher Scientific, San Jose, CA) equipped with an H-ESI probe and a Nanospace SI-2 HPLC system (Shiseido, Tokyo, Japan). Chromatographic separation was performed using a Hypersil GOLD column (150 x 2.1 mm, 3 µm; Thermo Fisher Scientific) at 40°C. The mobile phase consisted of acetonitrile-water (95:5, v/v) containing 0.1% acetic acid and was used at a flow rate of 200 µl/min. The general LC-MS/MS conditions were as follows: spray voltage, 1,000 V; vaporizer temperature, 350°C; sheath gas (nitrogen) pressure, 35 p.s.i.; auxiliary gas (nitrogen) flow, 45 arbitrary units; ion transfer capillary temperature, 350°C; collision gas (argon) pressure, 1.5 mTorr; collision energy, 16 V; and ion polarity, positive.

Evaluation of the solid-phase purification process
The absolute recovery of C4-7{alpha}-picolinate from the Bond Elut C18 cartridge was determined by the addition of [2H7]C4-7{alpha}-picolinate (1 ng) to the cartridge with a derivatizing mixture, which was prepared from 50 µl of serum without adding [2H7]C4. After the purification steps, the picolinoyl ester of 7ß-hydroxy-4-cholesten-3-one (1 ng) was added to the eluate as an internal standard, and both [2H7]C4-7{alpha}-picolinate and 7ß-hydroxy-4-cholesten-3-one-7ß-picolinate were quantified by LC-selected reaction monitoring (SRM).

To calculate the elimination rate of cholesterol by the Bond Elut C18 cartridge, unesterified cholesterol concentration in the first salting-out extraction was measured by colorimetric enzyme assay. The extract was then derivatized and purified by the cartridge, and the picolinoyl ester of [2H7]cholesterol (100 ng) was added to the final eluate as an internal standard. Cholesterol-3-picolinate and its 2H7 variant were quantified by the LC-SIM method to measure the absolute amount of cholesterol-3-picolinate contaminant in the final eluate.

Statistics
Data are reported as means ± SD. Reproducibility was analyzed by one-way layout (JMP software; SAS Institute, Inc., Cary, NC). Recovery was analyzed using a polynomial equation (20). Linearity of the calibration curve and correlation between serum volume and the amount of C4 were analyzed by simple linear regression. The regression analysis was also used to calculate the estimated amount ± 95% confidence limit in the recovery study. For all analyses, significance was accepted at P < 0.05.


    RESULTS
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
SRM
A typical ESI positive mass spectrum of the C4-7{alpha}-picolinate is shown in Fig. 1A . This picolinoyl ester derivative exhibited an [M+H]+ ion at m/z 506 as the base peak. In the MS/MS spectrum, the [M–C6H5O2N]+ ion was observed at m/z 383 as the most prominent peak, as shown in Fig. 1B. The SRM was conducted using m/z 506 -> m/z 383 for the C4-7{alpha}-picolinate and m/z 513 -> m/z 390 for the 2H7 variant.


Figure 1
View larger version (13K):
[in this window]
[in a new window]

 
Fig. 1. Typical electrospray ionization (ESI) positive mass spectrum (A) and product ion mass spectrum using m/z 506 as a precursor ion (B) of the picolinoyl ester derivative of 7{alpha}-hydroxy-4-cholesten-3-one (C4). Mass spectrometric conditions were as follows: introducing solvent, acetonitrile-water (95:5, v/v) containing 0.1% acetic acid; flow rate, 200 µl/min; spray voltage, 1,000 V; collision energy, 16 V.

 
Sensitivity of the present method
To determine the sensitivity of our LC-SRM method, the standard C4-7{alpha}-picolinate solution was diluted and injected into the LC-MS/MS system. As shown in Fig. 2A , the C4-7{alpha}-picolinate was easily detected to 100 fg, with a signal-to-noise ratio of 10, whereas conventional UV detection at 241 nm was barely able to detect 100 pg of C4 (Fig. 2B).


Figure 2
View larger version (19K):
[in this window]
[in a new window]

 
Fig. 2. Comparison of the detection limit of C4-7{alpha}-picolinate by selected reaction monitoring (SRM) at m/z 506 -> m/z 383 (A) with that of C4 by ultraviolet (UV) detection at 241 nm (B). Authentic standard of C4-7{alpha}-picolinate (100 fg) or C4 (100 pg) was injected into the HPLC system.

 
Absolute recovery of C4 and elimination rate of cholesterol in the solid-phase purification process
In this absolute recovery, [2H7]C4-7{alpha}-picolinate was used instead of the corresponding C4 derivative and was quantified by SRM using the picolinoyl ester of 7ß-hydroxy-4-cholesten-3-one as an internal standard. Both picolinoyl esters of C4 and 7ß-hydroxy-4-cholesten-3-one exhibited similar mass spectra, but the retention times were 4.66 and 5.00 min, respectively. Thus, the recovery of [2H7]C4-7{alpha}-picolinate was 104.0 ± 2.3% (n = 4).

The elimination rate of cholesterol by the Bond Elut C18 cartridge was calculated by quantifying cholesterol-3-picolinate contamination in the final eluate. The multiple-ion detector was focused on m/z 492 for cholesterol-3-picolinate and m/z 499 for [2H7]cholesterol-3-picolinate. Under the same HPLC conditions as for the determination of C4, the retention time of cholesterol-3-picolinate was ~11 min. The elimination rate of cholesterol by the cartridge was calculated as 98.6 ± 0.7% (n = 4).

Calibration curve
A calibration curve was established for C4 (Fig. 3 ). Different amounts (0.01, 0.05, 0.1, 0.25, 0.5, 1, 2, and 4 ng) of authentic C4 were mixed with 1 ng of [2H7]C4, derivatized to the picolinoyl ester, and purified by the Bond Elut C18 cartridge, as described in Materials and Methods. The weight ratio of C4, relative to the corresponding deuterated internal standard, was plotted on the abscissa, and the peak area ratio of the C4-7{alpha}-picolinate to the 2H7 variant, as measured by LC-SRM, was plotted on the ordinate. The linearity of the standard curve, as determined by simple linear regression, was excellent for weight ratios between 0.01 and 4.0 (n = 8; r = 1.000; P < 0.0001).


Figure 3
View larger version (10K):
[in this window]
[in a new window]

 
Fig. 3. Calibration curve for the weight ratio of C4 to the corresponding deuterated internal standard. Linearity was checked by simple linear regression, and the equation for the line of best fit was y = 1.035x + 0.008 (n = 8; r = 1.000; P < 0.0001).

 
Representative LC-SRM
Figure 4 shows typical LC-SRM chromatograms of C4-7{alpha}-picolinate and the 2H7 variant obtained from 50 µl of normal human serum. Both peaks corresponded to ~20 pg. The peak appearing at 5.46 min on this chromatogram (Fig. 4A) was identified as the picolinoyl ester derivative of 7-oxo-cholesterol by comparison with authentic 7-oxo-cholesterol-3-picolinate.


Figure 4
View larger version (14K):
[in this window]
[in a new window]

 
Fig. 4. Representative LC-SRM chromatograms of C4-7{alpha}-picolinate (A) and [2H7]C4-7{alpha}-picolinate (B) obtained from 50 µl of normal human serum. The peak of C4-7{alpha}-picolinate corresponds to ~20 pg (20 ng/ml).

 
Precision and accuracy of the present method
The following studies were performed to determine the precision and accuracy of our method using the same serum obtained from a normal human subject. Reproducibility was investigated by analyzing four samples in triplicate by LC-SRM (Tables 1 , 2 ). The results were analyzed by one-way layout, in which the analytical errors were divided into two sources of sample preparation and LC-SRM measurement. The variances were considered to be attributable to the measurement, because the errors during sample preparation were negligible. The interassay coefficients of variation for the between- and within-sample variations were 5.7% and 3.9%, respectively.


View this table:
[in this window]
[in a new window]

 
TABLE 1. Reproducibility in the quantification of C4 in human serum: analytical data

 

View this table:
[in this window]
[in a new window]

 
TABLE 2. Reproducibility in the quantification of C4 in human serum: ANOVA

 
For the recovery experiment (Table 3 ), known amounts of C4 (1.0–3.0 ng, 20.0–60.0 ng/ml) were spiked into 50 µl aliquots of the serum samples. After the clean-up procedure, LC-SRM was carried out in triplicate for each sample. The recoveries of the known spiked amounts of C4 ranged from 91.9% to 94.3%, with a mean of 93.4%. In addition, the amount of endogenous C4 found in unspiked 50 µl of serum was within the 95% confidence limit for the estimated amount of C4 calculated by linear regression analysis, which also constituted an index for the precision and accuracy of the method.


View this table:
[in this window]
[in a new window]

 
TABLE 3. Recovery of C4 from human serum

 
Application to a microscale assay
We studied the effect of reduced serum volume on the quantification of C4 amount. Different volumes (2, 5, 10, 25, and 50 µl) of normal human serum were diluted to 250 µl with distilled water, and extraction and quantification of C4 were conducted, as described in Materials and Methods. As shown in Fig. 5 , proportionality was observed when the volume of serum ranged from 2 µl to at least 50 µl for the amount of C4.


Figure 5
View larger version (9K):
[in this window]
[in a new window]

 
Fig. 5. Effects of reduced serum volume on quantification of the amount of C4. Linearity was checked by simple linear regression, and the equation for the line of best fit was y = 30.23x – 7.83 (n = 5; r = 0.999; P < 0.0001).

 

    DISCUSSION
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
HPLC with UV detection has been the most commonly used method for the determination of serum C4 concentrations (10, 11, 13, 21, 22). In HPLC-UV methods, a great deal of effort has been made to find appropriate internal standards and to develop superior clean-up procedures. In the latest method by Gälman et al. (22), 7ß-hydroxy-4-cholesten-3-one was added as an internal standard, and a commercially available solid-phase extraction column (C8) was used at a temperature of 64°C. One weak point of this method is that the sensitivity was not sufficient to quantify C4 in limited amounts of human serum. At least 1 ml of serum was needed for each assay, which is a considerably large volume for clinical blood chemistry tests.

We previously developed a more sensitive and specific method based on a stable isotope dilution technique using high-resolution GC-SIM (15). In this method, C4 was extracted from 200 µl of human plasma by a salting-out extraction, purified by a commercially available solid-phase extraction column (C18 and unbonded silica) at room temperature, and converted to the methyloxime-dimethylethylsilyl ether derivative before GC-SIM analysis. However, the sensitivity of this method was still not sufficient for the quantification of C4 in limited amounts of human serum, because the resulting derivatives exhibited two peaks of syn and anti isomers and their fragmentation patterns were not simple.

LC-MS or LC-MS/MS has come to be used more readily than GC-MS, because these methods do not always require a derivatization step. However, it is also true that the introduction of charged moieties enhances the ionization efficiency of neutral steroids in ESI and atmospheric pressure chemical ionization processes. A permanently charged N-methylpyridyl group was introduced into the hydroxyl group of cholesterol (23) and 5{alpha}-dihydrotestosterone (24, 25), whereas a permanently charged quaternary pyridinium moiety was introduced to the carbonyl group of oxosteroids with Girard P reagent (2628) or 2-hydrazino-1-methylpyridine (29). Very recently, Yamashita et al. (19) developed a new picolinoyl derivatization that is a simple and versatile method suitable for the sensitive and specific quantification of hydroxysteroids by positive ESI-LC-MS/MS. Interestingly, because the picolinoyl group is not charged permanently, even in the case of estradiol with two hydroxyl groups, a single charged ion was observed in the positive ESI mass spectrum. Therefore, picolinoyl ester may be a suitable derivative not only for monohydroxy steroids but also for hydroxycholesterols with plural hydroxyl groups.

Then, we applied this derivatization to quantify C4 in a very small volume of human serum by LC-SRM. The picolinoyl ester of C4 exhibited [M+H]+ as a single ion in ESI mode. The detection limit of C4 (as the C4 picolinoyl ester) in SRM mode was 100 fg, which was ~1,000 and 30 times more sensitive than that by the HPLC-UV and high-resolution GC-SIM methods, respectively. The detection limit of 100 fg corresponds to 0.1 ng/ml in 50 µl of serum, whereas the range of serum C4 concentrations has been reported to be 3–40 ng/ml (median, 12 ng/ml) in healthy subjects (10). Consequently, even when a sufficient volume of serum was not available, C4 concentrations could be determined by our method using a few microliters of serum. However, we recommend that 50 µl of serum be used for routine assay because some subjects treated with chenodeoxycholic acid (11) and patients with CYP7A1 deficiency (6) appear to exhibit unusually low serum concentrations of C4.

The derivatization and purification steps in this method are very simple. After the previously described salting-out extraction procedure (15), hydroxysteroids, including C4 and cholesterol, were derivatized to the picolinoyl esters within 30 min at room temperature. The C4-7{alpha}-picolinate was completely recovered from a commercially available C18 column with >98% elimination of cholesterol-3-picolinate. However, the amount of cholesterol-3-picolinate contaminant in the final eluate was still 300 times greater than that of C4-7{alpha}-picolinate. Therefore, the next injection into the LC-MS/MS system was not performed until cholesterol-3-picolinate had eluted completely from the column (~12 min), whereas the retention time of C4-7{alpha}-picolinate was ~4.7 min.

[2H7]C4 was added to serum as an ideal internal standard for quantification by SRM. Although the object of our study was to quantify a very small amount of C4, the specificity and reproducibility of this method were highly satisfactory.

In summary, a very sensitive and specific method for the quantification of C4 in human serum was developed. Derivatization of C4 into the picolinoyl ester made it possible to be quantified by LC-ESI-MS/MS with excellent sensitivity. Recovery and reproducibility experiments verified that this method provided analytical results with high reliability and reproducibility.


    ACKNOWLEDGMENTS
 
This work was supported in part by a grant from the Ministry of Education, Science, and Culture of Japan.

Manuscript received August 28, 2006 and in revised form October 31, 2006.


    REFERENCES
 TOP
 ABSTRACT
 INTRODUCTION
 MATERIALS AND METHODS
 RESULTS
 DISCUSSION
 REFERENCES
 
  1. Everson, G. T. 1992. Bile acid metabolism and its role in human cholesterol balance. Semin. Liver Dis. 12: 420–428.[Medline]

  2. Vlahcevic, Z. R., W. M. Pandak, D. M. Heuman, and P. B. Hylemon. 1992. Function and regulation of hydroxylases involved in the bile acid biosynthesis pathways. Semin. Liver Dis. 12: 403–419.[Medline]

  3. Vlahcevic, Z. R., R. T. Stravitz, D. M. Heuman, P. B. Hylemon, and W. M. Pandak. 1997. Quantitative estimations of the contribution of different bile acid pathways to total bile acid synthesis in the rat. Gastroenterology. 113: 1949–1957.[CrossRef][Medline]

  4. Schwarz, M., E. G. Lund, K. D. R. Setchell, H. J. Kayden, J. E. Zerwekh, I. Björkhem, J. Herz, and D. W. Russell. 1996. Disruption of cholesterol 7{alpha}-hydroxylase gene in mice. J. Biol. Chem. 271: 18024–18031.[Abstract/Free Full Text]

  5. Duane, W. C., and N. B. Javitt. 1999. 27-Hydroxycholesterol: production rates in normal human subjects. J. Lipid Res. 40: 1194–1199.[Abstract/Free Full Text]

  6. Pullinger, C. R., C. Eng, G. Salen, S. Shefer, A. K. Batta, S. K. Erickson, A. Verhagen, C. R. Rivera, S. J. Mulvihill, M. J. Malloy, et al. 2002. Human cholesterol 7{alpha}-hydroxylase (CYP7A1) deficiency has a hypercholesterolemic phenotype. J. Clin. Invest. 110: 109–117.[CrossRef][Medline]

  7. Root, C., C. D. Smith, S. S. Sundseth, H. M. Pink, J. G. Wilson, and M. C. Lewis. 2002. Ileal bile acid transporter inhibition, CYP7A1 induction, and antilipemic action of 264W94. J. Lipid Res. 43: 1320–1330.[Abstract/Free Full Text]

  8. Urizar, N. L., A. B. Liverman, D. T. Dodds, F. V. Silva, P. Ordentlich, Y. Yan, F. J. Gonzalez, R. A. Heyman, D. J. Mangelsdorf, and D. D. Moore. 2002. A natural product that lowers cholesterol as an antagonist ligand for FXR. Science. 296: 1703–1706.[Abstract/Free Full Text]

  9. Björkhem, I., E. Reihnér, B. Angelin, S. Ewerth, J-E. Åkerlund, and K. Einarsson. 1987. On the possible use of the serum level of 7{alpha}-hydroxycholesterol as a marker for increased activity of the cholesterol 7{alpha}-hydroxylase in humans. J. Lipid Res. 28: 889–894.[Abstract]

  10. Axelson, M., A. Aly, and J. Sjövall. 1988. Levels of 7{alpha}-hydroxy-4-cholesten-3-one in plasma reflect rates of bile acid synthesis in man. FEBS Lett. 239: 324–328.[CrossRef][Medline]

  11. Axelson, M., I. Björkhem, E. Reihnsér, and K. Einarsson. 1991. The plasma level of 7{alpha}-hydroxy-4-cholesten-3-one reflects the activity of hepatic cholesterol 7{alpha}-hydroxylase in man. FEBS Lett. 284: 216–218.[CrossRef][Medline]

  12. Hahn, C., C. Reichel, and K. von Bergmann. 1995. Serum concentration of 7{alpha}-hydroxycholesterol as an indicator of bile acid synthesis in humans. J. Lipid Res. 36: 2059–2066.[Abstract]

  13. Sauter, G., F. Berr, U. Beuers, S. Fischer, and G. Paumgartner. 1996. Serum concentrations of 7{alpha}-hydroxy-4-cholesten-3-one reflect bile acid synthesis in humans. Hepatology. 24: 123–126.

  14. Yoshida, T., A. Honda, N. Tanaka, Y. Matsuzaki, B-F. He, T. Osuga, N. Kobayashi, K. Ozawa, and H. Miyazaki. 1993. Simultaneous determination of mevalonate and 7{alpha}-hydroxycholesterol in human plasma by gas chromatography-mass spectrometry as indices of cholesterol and bile acid biosynthesis. J. Chromatogr. 613: 185–193.[Medline]

  15. Yoshida, T., A. Honda, N. Tanaka, Y. Matsuzaki, J. Shoda, B-F. He, T. Osuga, and H. Miyazaki. 1994. Determination of 7{alpha}-hydroxy-4-cholesten-3-one level in plasma using isotope-dilution mass spectrometry and monitoring its circadian rhythm in human as an index of bile acid biosynthesis. J. Chromatogr. 655: 179–187.

  16. Honda, A., T. Yoshida, G. Xu, Y. Matsuzaki, S. Fukushima, N. Tanaka, M. Doy, S. Shefer, and G. Salen. 2004. Significance of plasma 7{alpha}-hydroxy-4-cholesten-3-one and 27-hydroxycholesterol concentrations as markers for hepatic bile acid synthesis in cholesterol-fed rabbits. Metabolism. 53: 42–48.[CrossRef][Medline]

  17. Honda, A., J. Shoda, N. Tanaka, Y. Matsuzaki, T. Osuga, N. Shigematsu, M. Tohma, and H. Miyazaki. 1991. Simultaneous assay of the activities of two key enzymes in cholesterol metabolism by gas chromatography-mass spectrometry. J. Chromatogr. 565: 53–66.[Medline]

  18. Shimasue, A. 1974. The action of cholesterol:oxygen oxidoreductase on cholest-5-ene-3ß,7{alpha}-diol and the enzymatic preparation of the labeled cholest-4-en-7{alpha}-ol-3-one. Hiroshima J. Med. Sci. 23: 265–272.[Medline]

  19. Yamashita, K., S. Kobayashi, S. Tsukamoto, and M. Numazawa. 2006. Synthesis of pyridine-carboxylate derivatives of hydroxysteroids for liquid chromatography-electrospray ionization-mass spectrometry. Steroids. Epub ahead of print. November 30, 2006; doi: 10.1016/j. steroids.2006.10.005.

  20. Taguchi, G. 1986. Introduction to Quality Engineering—Designing Quality into Products and Process. Asian Productivity Organization, Tokyo, Japan.

  21. Pettersson, L., and C. G. Eriksson. 1994. Reversed-phase high-performance liquid chromatographic determination of 7{alpha}-hydroxy-4-cholesten-3-one in human serum. J. Chromatogr. 657: 31–36.[Medline]

  22. Gälman, C., I. Arvidsson, B. Angelin, and M. Rudling. 2003. Monitoring hepatic cholesterol 7{alpha}-hydroxylase activity by assay of the stable bile acid intermediate 7{alpha}-hydroxy-4-cholesten-3-one in peripheral blood. J. Lipid Res. 44: 859–865.[Abstract/Free Full Text]

  23. Quirke, J. M. E., C. L. Adams, and G. J. Van Berkel. 1994. Chemical derivatization for electrospray ionization mass spectrometry. I. Alkyl halides, alcohols, phenols, thiols, and amines. Anal. Chem. 66: 1302–1315.

  24. Nakagawa, Y., and Y. Hashimoto. 2002. Polar derivatization of 5{alpha}-dihydrotestosterone and sensitive analysis by semimicro-LC/ESI-MS. J. Mass Spectrom. Soc. Jpn. 50: 330–336.

  25. Nishiyama, T., Y. Hashimoto, and K. Takahashi. 2004. The influence of androgen deprivation therapy on dihydrotestosterone levels in the prostatic tissue of patients with prostate cancer. Clin. Cancer Res. 10: 7121–7126.[Abstract/Free Full Text]

  26. Lai, C. C., C. H. Tsai, F. J. Tsai, C. C. Lee, and W. D. Lin. 2001. Rapid monitoring assay of congenital adrenal hyperplasia with microbore high-performance liquid chromatography/electrospray ionization tandem mass spectrometry from dried blood spots. Rapid Commun. Mass Spectrom. 15: 2145–2151.[CrossRef][Medline]

  27. Griffiths, W. J., S. Liu, G. Alvelius, and J. Sjovall. 2003. Derivatisation for the characterisation of neutral oxosteroids by electrospray and matrix-assisted laser desorption/ionisation tandem mass spectrometry: the Girard P derivative. Rapid Commun. Mass Spectrom. 17: 924–935.[CrossRef][Medline]

  28. Griffiths, W. J., Y. Wang, G. Alvelius, S. Liu, K. Bodin, and J. Sjövall. 2006. Analysis of oxysterols by electrospray tandem mass spectrometry. J. Am. Soc. Mass Spectrom. 17: 341–362.[CrossRef][Medline]

  29. Higashi, T., A. Yamauchi, and K. Shimada. 2005. 2-Hydrazino-1-methylpyridine: a highly sensitive derivatization reagent for oxosteroids in liquid chromatography-electrospray ionization-mass spectrometry. J. Chromatogr. B Analyt. Technol. Biomed. Life Sci. 825: 214–222.[Medline]


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
J. Lipid Res.Home page
A. Honda, K. Yamashita, H. Miyazaki, M. Shirai, T. Ikegami, G. Xu, M. Numazawa, T. Hara, and Y. Matsuzaki
Highly sensitive analysis of sterol profiles in human serum by LC-ESI-MS/MS
J. Lipid Res., September 1, 2008; 49(9): 2063 - 2073.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
M. Lenicek, M. Juklova, J. Zelenka, J. Kovar, M. Lukas, M. Bortlik, and L. Vitek
Improved HPLC Analysis of Serum 7{alpha}-Hydroxycholest-4-en-3-one, a Marker of Bile Acid Malabsorption
Clin. Chem., June 1, 2008; 54(6): 1087 - 1088.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
D600032-JLR200v1
48/2/458    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Honda, A.
Right arrow Articles by Miyazaki, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Honda, A.
Right arrow Articles by Miyazaki, H.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 All ASBMB Journals   Journal of Biological Chemistry 
 Molecular and Cellular Proteomics   ASBMB Today