|
|
||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Methods |
Department of Cardiology, Fukuoka University School of Medicine, Fukuoka, Japan
Published, JLR Papers in Press, January 10, 2007.
1 To whom correspondence should be addressed. e-mail: bozhang{at}fukuoka-u.ac.jp
| ABSTRACT |
|---|
|
|
|---|
(iPF2
-III) when the methanol concentration was increased under acidic, neutral, and base wash conditions. A multidimensional (MD)-SPE method that incorporated size exclusion, reverse-phase chromatography, and normal-phase chromatography was developed using an Oasis HLB SPE cartridge and an HLB µElution SPE plate. Average extraction recoveries of the deuterated internal standards of iPF2
-III and iPF2
-VI were 62 ± 8% and 60 ± 10%. A buffer-free HPLC method for the separation of F2-iP isomers was developed on base-deactivated C8 columns. Average matrix effects for iPF2
-III and iPF2
-VI were 95 ± 6% and 103 ± 5%. The clean extraction of urine F2-iPs using MD-SPE and the separation of F2-iP isomers using a novel HPLC method did not cause apparent ion suppression in the analysis of iPF2
-III and iPF2
-VI using LC-MS/MS. These findings should be useful for establishing a routine LC-MS/MS method for the analysis of F2-iPs.
Supplementary key words liquid chromatography-tandem mass spectrometry urine ion suppression effects buffer-free HPLC sample preparation iPF2
-VI
| INTRODUCTION |
|---|
|
|
|---|
The measurement of F2-iPs in biological samples presents several challenges. First, the methods used for measurement have to be specific, because F2-iPs are isomers of F2-PGs and exist as four sets of regioisomers (1, 11). In addition, the methods used for measurement have to be sensitive, because F2-iPs exist at very low concentrations in biological samples (1215).
F2-iPs are usually measured noninvasively in urine. The 8-iso-PGF2
[8-iso-15(S) PGF2
, also iPF2
-III] is the F2-iP that has been studied most often. iPF2
-VI, a regioisomer of iPF2
-III, has been shown to be the more abundant F2-iP (16). GC-MS methods are the "gold standard" for the measurement of F2-iPs (13, 17, 18). GC-MS methods are highly sensitive but also laborious and are not sufficiently selective (13, 18). ELISA kits for measuring F2-iPs are commercially available, but the simultaneous analysis of F2-iP regioisomers is not possible. Selective and simultaneous measurements of different F2-iPs could be important because they may have different biological activities and may be formed and metabolized differently under disease conditions that are linked to oxidative stress (19).
Liquid chromatography-tandem mass spectrometry (LC-MS/MS) methods have been developed for the measurement of F2-iPs and their metabolites (1, 14, 15, 20). LC-MS/MS methods are specific, because HPLC can separate F2-iP stereoisomers and electrospray ionization (ESI)-MS/MS can separate F2-iP regioisomers. Also, LC-MS/MS methods are sensitive because the ESI-MS/MS technique permits the detection of very low levels of target analytes in the presence of a complex matrix background. However, LC-MS/MS is also associated with a major drawback when analyzing target analytes in biological fluids: the electrospray source is very susceptible to matrix-related ion suppression effects (21), and these matrix effects may vary from sample to sample (22).
Stable isotope-labeled analogs, which are chemically and structurally the same as the target analytes but differ in molecular mass, have been used as internal standards (IS) to compensate for variations in injection, sample preparation, instrument parameter, and matrix effects. However, interference compounds that coelute with deuterated IS of 8-iso-PGF2
have been shown to be contained in plasma samples prepared by solvent extraction (23). This gives inaccurate results, because target analytes and IS ratios are used to calculate analyte concentrations. This problem can be detected and circumvented by prolonged gradient HPLC separation (23). However, matrix-related compounds that coelute with the target analytes may also give inaccurate results, and these problems cannot be detected. Therefore, the clean extraction of biological samples is essential for the accurate and reproducible analysis of F2-iPs using LC-MS/MS.
Matrix-related ion suppression effects have not been examined in previously reported LC-MS/MS methods for the analysis of F2-iPs in biological samples (14, 15, 23, 24). Because the control of matrix effects is important for establishing LC-MS/MS as a routine technique for the measurement of F2-iPs in biological samples, we developed a novel urine sample clean-up technique using multidimensional (MD)-solid-phase extraction (SPE) and a novel buffer-free HPLC separation method to minimize matrix- and buffer-additive-related ion suppression effects in the analysis of F2-iPs using LC-MS/MS.
| METHODS |
|---|
|
|
|---|
-III, (±)5-iPF2
-VI, and PGF2
shown in Scheme 1
were obtained from Cayman Chemical (Ann Arbor, MI). Captiva Filter cartridges (0.2 µm) were obtained from Varian, Inc. (Palo Alto, CA). Oasis HLB (hydrophilic-lipophilic balance) SPE cartridges, Oasis MAX (mixed-mode anion exchange) SPE cartridges, Oasis MCX (mixed-mode cation exchange) SPE cartridges, and Oasis HLB µElution SPE plates were purchased from Waters Corp. (Milford, MA). Strata X SPE cartridges were obtained from Phenomenex (Torrance, CA). Methanol (MeOH), acetonitrile (ACN), and water were LC-MS grade (Wako Pure Chemical Industries, Ltd., Osaka, Japan). All other reagents were HPLC grade.
|
standard solution with a microsyringe pump (10 µl/min). The optimum tuning conditions for ESI were as follows: capillary voltage, 3 kV; source block temperature, 120°C; extractor, 3.0 V; radio frequency (RF) lens, 0.1 V; desolvation gas (nitrogen) heated to 400°C and delivered at a flow rate of 1,200 l/h; cone gas flow, 50 l/h; ion energy, 1.0; multiplier, 650 V; and low and high mass resolutions, 13. Entrance, collision, and exit for MS tuning were 50, 2, and 50 V, respectively, and entrance, exit, and collision gas (argon gas) flow for MS/MS tuning were 2 V, 2 V, and 0.35 ml/min, respectively. A fully automated instrument mass calibration (static calibration, scanning calibration, and scan speed compensation) was performed using MassLynx version 4.1 software in the ES+ mode with MS tuning parameters (capillary voltage, 3 kV; cone voltage, 40 V; source block temperature, 80°C; extractor, 3.0 V; RF lens, 0.1 V; and desolvation gas heated to 150°C and delivered at a flow rate of 350 l/h) and the NAICS calibration reference file. Waters Atmospheric Pressure Ionization Calibration solution (NaCsI; catalog No. 700001593) was introduced using the instrument's syringe pump (250 µl) at a pump flow of 10 µl/min.
Nominal mass data acquisition parameters for 2,3-dinor-8-iso-PGF2
(2,3-dinor-iPF2
-III), iPF2
-III, iPF2
-III-d4, iPF2
-VI, and iPF2
-VI-d11 were determined using flow injection (5 µl, 0.2 ml/min) (without the HPLC column) of standard solutions of the individual compounds (1,000 ng/ml). The mass-to-charge ratio (m/z) of the precursor ion of an individual compound was determined in the MS scan mode, and the sample cone voltage for maximum signal intensity of the deprotonated species [M-H] was determined in the selected-ion reaction mode. Collision-induced dissociation of each deprotonated molecule was performed using MS/MS tuning. The most abundant daughter ion of F2-iPs and deuterated IS was determined in the daughter scan mode, and the collision energy for maximum signal intensity of precursor-product ions was determined in the multiple reaction monitoring (MRM) mode. An interchannel delay time of 0.01 s, interscan delay time of 0.01 s, and dwell time of 0.1 s were used in data acquisition. The optimized cone voltage and collision energy for the individual compounds are given in Fig. 1
.
|
Development of a selective SPE procedure
For the efficient development of a selective SPE procedure for the extraction of F2-iPs in urine samples, [3H]8-iso-PGF2
(
10,000 cpm/ml; Cayman Chemical, Ann Arbor, MI) was added to acidified urine samples and extracted together with 8-iso-PGF2
(3 ng/ml). Oasis SPE cartridges (3 cc/60 mg) were conditioned with 2 ml of MeOH and 2% formic acid, and SPE extraction was performed under vacuum using a Vac Elut SPS 24 Manifold (Varian, Inc.). The retention capacities of different Oasis SPE sorbents (i.e., HLB, MAX, and MCX) were examined by applying urine samples (1 ml each up to 10 ml) to the respective SPE cartridges. Two hundred microliters of wash waste and eluate were collected and counted for radioactivity in 3 ml of liquid scintillation solution (Clear-sol I; Nacalai Tesque, Inc., Kyoto, Japan) on a liquid scintillation counter (Tri-carb 2500; Packard Instrument Co., Meriden, CT).
A selective SPE wash procedure was developed by taking advantage of the differences in the elution profiles of F2-iPs and sample matrix interference as a function of both the pH of the wash and the concentration of the elution solvents (27). The percentage of MeOH in acidic, neutral, and base washes was determined by washing the cartridges with 2% formic acid, water, and 2% NH4OH solutions containing varying concentrations (10100%) of MeOH, respectively, and collecting the waste for radioactivity counting. Elution volume was determined by eluting F2-iPs with six 0.5 ml portions of diethyl ether-acetic acid (100:2) solvent and counting the radioactivity in the eluate.
Development of a MD-SPE method
An Oasis HLB cartridge (1 cc/30 mg) and an Oasis HLB µElution plate (750 µl/2 mg) were used to develop the MD-SPE method.
Two milliliters of centrifuged urine samples was mixed with one-fifth volume of MeOH and mixtures of IS solutions containing 8-iso-PGF2
-d4, PGF2
-d4, and (±)5-iPF2
-VI-d11 and put on ice for 30 min. The samples were adjusted to pH
3 with 12 ml of 1% formic acid before extraction. The HLB SPE cartridges were conditioned with 1 ml of MeOH and 1 ml of 5% MeOH/2% formic acid solution before use. The acidified and diluted urine samples were applied to a 0.2 µm Captiva Filter cartridge that was attached to the top of an Oasis HLB SPE cartridge using a Bond Elute adaptor (Varian, Inc.) with a 20 Port Vacuum Manifold (Agilent Technologies, Palo Alto, CA) at a flow rate of <3 ml/min. After sample application, filter cartridges were discarded and the SPE cartridge was washed with 1 ml of 5% MeOH. Subsequent wash and elution steps were continued using a Cerex System 48 Positive-Pressure Manifold (SPEware Corp., San Pedro, CA). Nitrogen gas was used for positive-pressure SPE.
Eluate from Oasis HLB SPE cartridges was applied to an Oasis µElution plate. SPE extraction with an Oasis µElution plate was performed using a Plate SPE Extraction Manifold (Waters Corp.). Eluate from the Oasis µElution plate was diluted with water and analyzed directly using LC-MS/MS.
Assessment of extraction recovery and matrix-related ion suppression effects
Recovery in urine sample extraction was examined by adding the IS mixtures [8-iso-PGF2
-d4 and (±)5-iPF2
-VI-d11] to urine samples before and after extraction. Recovery (%) was calculated as peak area of IS spiked into urine samples before extraction/peak area of IS spiked into urine samples after extraction x 100 (28).
Ion suppression effects of sample matrices were examined by adding mixtures of standard solutions [2,3-dinor-8-iso-PGF2
, 15(R)-8-iso-PGF2
, 8-iso-PGF2
, and (±)5-iPF2
-VI] and deuterated IS solutions [8-iso-PGF2
-d4 and (±)5-iPF2
-VI-d11] to water and sample matrix, respectively. The matrix effect (%) was calculated as peak area of the standard or IS added into urine samples/peak area of the standard or IS added into water x 100 (28).
Data analysis
Peak-to-peak signal-to-noise ratios for iPF2
-III and iPF2
-VI peaks on mass chromatograms were determined using MassLynx version 4.1 software after smoothing using the Moving Mean method [window size (scans) set at ±3, number of smoothings set at 2]. Significant differences in the signal-to-noise ratios of iPF2
-III or iPF2
-VI peaks determined from 1 and 2 ml urine samples were determined by the Wilcoxon sign-rank test using the SAS (Statistical Analysis System) software package (version 9.1; SAS Institute, Inc., Cary, NC) at Fukuoka University. QuanLynx software (Waters Corp.) was used for automated peak identification and the integration of peak area on the MRM chromatogram. The integrated area data for the analyzed samples were exported as text files and combined using a self-made Excel macro. Recovery and matrix effects were calculated using the SAS software package.
| RESULTS |
|---|
|
|
|---|
; G, 8-iso-PGF2ß; J, 8-iso-15(R) PGF2
], group VI F2-iPs [L, (±)5-iPF2
-VI], F2-PG [B, PGF2
; C, 11ß-PGF2
; F, 5-trans-PGF2
; H, PGF2ß; I, 5-trans-PGF2ß; K, 15(R) PGF2
], and metabolites of 8-iso-PGF2
(M, 2,3-dinor-8-iso-PGF2
) and PGF2
(N, 2,3-dinor-PGF2
). Scheme 1 also shows the structures of deuterated IS of iPF2
-III (D, 8-iso-PGF2
-d4), PGF2
(E, PGF2
-d4), and iPF2
-VI [O, (±)5-iPF2
-VI-d11].
Figure 1 shows the product ion spectra of F2-iPs. Abundant molecular ions generated under the ES-mode were m/z 353 for 8-iso-PGF2
(Fig. 1A) and (±)5-iPF2
-VI (Fig. 1D), m/z 357 for 8-iso-PGF2
-d4 (Fig. 1B), m/z 325 for 2,3-dinor-8-iso-PGF2
(Fig. 1C), and m/z 364 for (±)5-iPF2
-VI-d11 (Fig. 1E). Product ions at m/z 193 (Fig. 1A) and m/z 115 (Fig. 1D) were the most abundant product ions and were specific for 8-iso-PGF2
and iPF2
-VI, respectively. The most abundant product ions were m/z 197 for 8-iso-PGF2
-d4 (Fig. 1B), m/z 237 for 2,3-dinor-8-iso-PGF2
(Fig. 1C), and m/z 115 for iPF2
-VI-d11 (Fig. 1E). Therefore, precursor/product ion pairs for the detection of F2-iPs under MRM mode were determined to be as follows: 8-iso-PGF2
m/z 353/193 for 8-iso-PGF2
, m/z 357/197 for 8-iso-PGF2
-d4, m/z 325/237 for 2,3-dinor-8-iso-PGF2
, m/z 353/115 for iPF2
-VI, and m/z 364/115 for iPF2
-VI-d11.
8-Iso-PGF2
and PGF2
showed similar fragmentation patterns in the product ion spectra (data not shown). Therefore, the separation of 8-iso-PGF2
from PGF2
isomers is necessary for the specific determination of 8-iso-PGF2
.
Development of a reverse-phase HPLC method for the separation of F2-iP isomers
As a first step in the development of a novel HPLC method, experiments were performed to determine the composition of the mobile phase using an ACE 3 µm C8 50 x 2.1 mm i.d. column. Because we found that the ionization efficiency of 8-iso-PGF2
was higher in MeOH than in ACN (data not shown), the separation of isomers of F2-iPs began with MeOH, and the proportion of ACN was increased to increase the selectivity. The resolution of four isomers, 15(R) iPF2
-III, iPF2
-III, 15(R) PGF2
, and PGF2
, was increased when the proportion of ACN to MeOH was increased, and enough separation was obtained when the proportion of ACN to MeOH was increased to 1:2 (data not shown).
Because ammonium acetate is the common buffer additive used in the analysis of 8-iso-PGF2
by LC-MS/MS (1, 14, 15), we examined the effects of ammonium acetate on the ionization efficiency of 15(R) iPF2
-III, iPF2
-III, 15(R) PGF2
, and PGF2
by adding ammonium acetate solutions (pH 6.8) to the standard solutions. Surprisingly, the ion intensities of all four isomers were greatly reduced when even low concentrations (0.4 and 2 mM) of ammonium acetate were present in the standard solutions (data not shown). Ammonium acetate decreased the retention of all four isomers but did not affect the separation (data not shown).
The effects of pH on the separation of isomers were examined by analyzing mixtures of four standard solutions prepared with water and 0.001, 0.01, and 0.1% acetic acid, respectively. Retention of the four isomers increased with an increase in the concentration of acetic acid, but separation was not affected. Therefore, the following HPLC method was developed under neutral pH without buffer additives.
The optimum temperature and gradient conditions for separation of the four isomers were determined using the computer simulation software DryLab 2000 Plus. Four experiments were performed with simultaneous changes in temperature (20°C and 40°C) and gradient steepness [gradient time, 6 and 18 min; gradient, 4090% MeOH/ACN (2:1)]. The computer simulation indicated that better separation can be achieved under a shallow gradient condition and at lower temperature.
Finally, to optimize peak shape and separation under neutral conditions, C8 columns from different makers were screened for the separation of standard solutions of six PGF2
isomers: 8-iso-15(R) PGF2
, 8-iso-PGF2
, 11ß-PGF2
, 15(R) PGF2
, 5-trans-PGF2
, and PGF2
. A Hypersil BDS 3 µm C8 50 x 2.1 mm i.d. column was chosen as the analysis column based on the symmetric peak, good separation, and slightly stronger retention of F2-iP isomers than with the ACE C8 column (data not shown).
Figure 2
shows LC-MS/MS results with the standard solutions of F2-iPs and F2-PGs. As shown in Fig. 2A, iPF2
-III (8-iso-PGF2
) was baseline-separated from 15(R) iPF2
-III [8-iso-15(R) PGF2
], 11ß-PGF2
, 15(R) PGF2
, 5-trans-PGF2
, and PGF2
. As shown in Fig. 2B, 8-iso-PGF2ß and PGF2ß were coeluted with iPF2
-III and 5-trans-PGF2ß was coeluted with 11ß-PGF2
. Other stereoisomer pairs (Fig. 2CF), iPF2
-III-d4 (8-iso-PGF2
-d4) and PGF2
-d4, 2,3-dinor-8-iso-PGF2
and 2,3-dinor-PGF2
, iPF2
-VI and 5-epi-iPF2
-VI, and iPF2
-VI-d11 and 5-epi-iPF2
-VI-d11, were all baseline-separated from each other. As also shown in Fig. 2, not only baseline separation but also sharp and symmetric peaks were achieved using the Hypersil BDS 3 µm C8 column.
|
added to a urine sample for Oasis HLB, MAX, and MCX SPE cartridges (3 cc/60 mg). As shown, [3H]8-iso-PGF2
in up to 10 ml urine sample was consistently retained by Oasis HLB cartridges but not by Oasis MAX or MCX cartridges. This result indicates that Oasis HLB had the greatest retention capacity; therefore, it was used in the following SPE method development.
|
when the percentage of MeOH was increased from 0% in 10% increments at acidic pH (with 2% formic acid), neutral pH (with water), and alkaline pH (with 2% NH4OH) washes. As shown in Fig. 3B, [3H]8-iso-PGF2
was retained until the content of MeOH was increased to 50% under acidic and neutral wash conditions, but it was only retained when the content of MeOH was <20% under basic wash conditions. This result indicates that the elution profile of [3H]8-iso-PGF2
as a function of MeOH concentration was different at low and high pH. According to these results, a selective SPE procedure was developed and incorporated into MD-SPE.
MD-SPE of urine samples for LC-MS/MS analysis
Figure 3C shows the MD-SPE method developed for the extraction of F2-iPs in urine. First, an acidified (pH
3) urine sample containing 10% MeOH was applied to an HLB cartridge and washed with 5% MeOH. In this sample application and cartridge wash step 1 (Fig. 3C), high-molecular-weight sample components (such as proteins) were size-excluded because of the small pore diameter (8 nm) of the sorbent, and salts and other polar interference such as carbohydrates were not retained by reverse-phase chromatography. In cartridge wash step 2 (base wash), acidic, moderately polar, hydrophobic interference was washed off using 5% MeOH with 2% NH4OH (Fig. 3C). The elution of yellow interference from urine samples was observed in this step. In cartridge wash step 3, pH was switched to acidic by washing with 5% MeOH and 2% formic acid (Fig. 3C). In cartridge wash step 4, basic, moderately polar, hydrophobic interference was washed off using 15% MeOH with 2% formic acid. In cartridge wash steps 5 and 6, neutral interference was washed off with hexane-diethyl ether (9:1) and hexane-ethyl acetate (9:1), respectively (Fig. 3C). In elution fraction step A, the fatty acid fraction was selectively eluted with diethyl ether with 2% acetic acid (Fig. 3C). Visible yellow interference remained on the SPE cartridges.
Eluted fraction A from the HLB SPE cartridge was applied to the HLB µElution plate after dilution with hexane (Fig. 3C). In this step, F2-iPs were retained on the SPE sorbent by normal-phase chromatography. In plate wash step 1, acetic acid was washed off with 5% MeOH (Fig. 3C). In plate wash step 2 (base wash), yellow interference was washed off with 5% MeOH and 2% NH4OH. In plate wash steps 3 and 4, the pH was switched to acidic by washing with 5% MeOH and 2% formic acid, and formic acid was subsequently washed off with 5% MeOH (Fig. 3C). In plate wash step 5, moderately polar, hydrophobic interference was washed off with 15% MeOH. Finally, in elution fraction step B, F2-iP fractions were eluted with 80% MeOH. The final eluted fraction was clear and used directly for LC-MS/MS analysis after dilution with water.
Analysis of urine extracts using LC-MS/MS
Figure 4
shows MD-SPE LC-MS/MS results for a urine sample detected under MRM mode. As shown in Fig. 4A, 8-iso-15(R) PGF2
and 8-iso-PGF2
were baseline-separated from PGF2
and other unknown isomers. 8-Iso-15(R) PGF2
, 8-iso-PGF2
, 15(R) PGF2
, and PGF2
were identified by adding mixtures of the standard solutions to the extracted sample (Fig. 4B). Figure 4C shows deuterium-labeled 8-iso-PGF2
(8-iso-PGF2
-d4) and PGF2
(PGF2
-d4) extracted from the urine sample. Extraction of blank sample (not spiked with deuterated IS) indicated that no interference was coeluted with 8-iso-PGF2
-d4 and PGF2
-d4 (Fig. 4D). Figure 4E shows that 2,3-dinor-iPF2
-III was baseline-separated from other unknown isomers. Increasing the gradient time and/or the use of a longer column baseline-separated other unknown isomers (data not shown). As shown in Fig. 4F, iPF2
-VI and 5-epi-iPF2
-VI were baseline-separated from other unknown isomers. Figure 4G shows that iPF2
-VI-d11 and 5-epi-iPF2
-VI-d11 extracted from the urine sample were baseline-separated from each other. No interference was coeluted with iPF2
-VI-d11 and 5-epi-iPF2
-VI-d11, as demonstrated by extraction of a blank sample (data not shown).
|
-III and iPF2
-VI were quantitatively recovered from different volumes of urine samples, 1 and 2 ml urine samples from eight volunteer subjects were spiked with the same amount of iPF2
-III-d4 and iPF2
-VI-d11 (2 ng) and extracted using the MD-SPE method. Figure 5
shows that the relative areas of iPF2
-III to iPF2
-III-d4 (Fig. 5A) and iPF2
-VI to iPF2
-VI-d11 (Fig. 5B) in 2 ml urine samples were well correlated (r = 0.992 and 0.997, respectively) with those in 1 ml urine samples. The signal-to-noise ratios for iPF2
-III (Fig. 5C) and iPF2
-VI (Fig. 5D) ranged from 16 to 118 and from 143 to 355, respectively. As shown in Fig. 5, the signal-to-noise ratios for iPF2
-III (Fig. 5C) and iPF2
-VI (Fig. 5D) extracted from 2 ml urine samples were greater than those extracted from 1 ml urine samples. These results indicate that peak-to-peak signal-to-noise ratios of >20 were achieved for iPF2
-III and iPF2
-VI analysis by extracting 2 ml urine samples from eight randomly selected volunteers.
|
-d4 and (±)5-iPF2
-VI-d11 (2 ng) were added to 1 or 2 ml urine samples before and after extraction to examine sample extraction recovery (Table 1
). As shown in Table 1 (extraction 1),
5070% of the IS was recovered from 1 and 2 ml urine samples from four volunteer subjects. Four repetitive extractions (extractions 14) of 2 ml urine samples from four subjects showed consistent recoveries, and the average recoveries for iPF2
-III-d4, iPF2
-VI-d11, and 5-epi-iPF2
-VI-d11 were 62 ± 8%, 60 ± 10%, and 61 ± 8%, respectively (Table 1).
|
-III, which was extracted from 1 and 2 ml urine samples (extraction 1) or at different times (extractions 14). The matrix effects for 1 and 2 ml urine samples and different extractions on other F2-iPs ranged from 88% to 127% (Table 2), and the average matrix effects for 15(R) iPF2
-III, iPF2
-III, iPF2
-III-d4, iPF2
-VI, 5-epi-iPF2
-VI, iPF2
-VI-d11, and 5-epi-iPF2
-VI-d11 were 95 ± 6%, 100 ± 6%, 103 ± 5%, 96 ± 3%, 100 ± 3%, 109 ± 15%, and 104 ± 16%, respectively. These results indicate that the ion suppression effects of urine matrices did not significantly affect the ionization efficiency of these F2-iPs.
|
| DISCUSSION |
|---|
|
|
|---|
In this work, we developed a novel MD-SPE and buffer-free HPLC method to control matrix- and buffer-additive-related ion suppression in the LC-MS/MS analysis of F2-iPs in urine samples.
We developed the MD-SPE method based on two-step SPE, applying orthogonal retention mechanisms, and on a selective SPE wash and elution protocol for F2-iPs. In the first-step SPE on Oasis HLB SPE cartridges, F2-iPs were retained by a reverse-phase chromatography mechanism, and in the second-step SPE on an Oasis µElution SPE plate, F2-iPs were retained by a normal-phase chromatography mechanism. Because Oasis HLB is a hydrophilic-lipophilic balanced polymer sorbent (27), both reverse-phase chromatography and normal-phase chromatography mechanisms can be used for the retention of F2-iPs. The solvents were exchanged for LC-MS/MS analysis in the second-step SPE, which eliminated the need to evaporate organic solvents that is normally part of the SPE of urine samples (14, 15, 23, 24). Also, the samples were concentrated using small-bed SPE sorbent (2 mg) in the second-step SPE (Fig. 3C). Concentration is normally achieved by the evaporation of organic solvent, and both target analytes and matrix interference are concentrated. However, in the second-step SPE on the Oasis µElution plate, F2-iPs were concentrated but matrix interference was removed, which further cleaned up the samples. Because of the small elution volume (40 µl), a 10-fold concentration of 2 ml urine samples was easily achieved by the MD-SPE method. In addition, using MD-SPE, the time required to prepare samples was greatly shortened, because organic solvents did not have to be evaporated.
In MD-SPE, the clean extraction of urine samples was achieved by a novel selective SPE wash and elution procedure. Oasis HLB is a polymer sorbent and can be used from pH 1 to 14 (27). Therefore, a selective SPE wash procedure was developed by taking advantage of the different elution profiles of F2-iPs and urine matrix interference as a function of both the concentration of MeOH and pH. With HLB sorbent, acid and base compounds show opposite retention behaviors under low and high pH. Under low pH, acid compounds have strong retention and base compounds have weak retention, whereas under high pH, acid compounds have weak retention and base compounds have strong retention. Therefore, when acidified urine samples containing 10% MeOH were applied to an HLB SPE cartridge, base interference was not retained. Acidic wash with 15% MeOH/2% formic acid (Fig. 3C, cartridge wash step 4) was also used to eliminate base interference. Base wash with 5% MeOH/2% NH4OH was used to remove acid interference. Elution of yellow interference was observed during the base wash.
Neutral interference in the urine samples was washed off with hexane. Because hexane is immiscible with MeOH and water, one part of diethyl ether was added to the wash solvent (Fig. 3C, cartridge wash step 5). A wash step with hexane-ethyl acetate (9:1) (Fig. 3C, cartridge wash step 6) was further included to remove residual water on the HLB SPE cartridge because ethyl acetate is miscible with water. Diethyl ether/2% acetic acid was used for the elution of F2-iPs in the first-step SPE because diethyl ether is less polar and more selective than ethyl acetate. Therefore, selective wash and selective elution gave a selective SPE procedure for the clean extraction of F2-iPs.
The cleanliness of the extracted urine samples as examined by MS was not markedly affected by slight variations in extraction parameters, including the concentration of organic solvent and pH (data not shown), indicating that the selective SPE procedure was robust. Also, the use of Strata X SPE cartridges, another polymer sorbent, instead of Oasis HLB SPE cartridges did not affect the cleanliness of extraction or recovery (data not shown), indicating that the MD-SPE method is robust.
Because F2-iPs are isomers of F2-PGs and consist of many diastereomers (1, 11), separation of these isomers is important for the specific analysis of individual F2-iPs. Li et al. (15) showed that 15(R) 8-iso-PGF2
and 8-iso-PGF2
were baseline-separated using Hypersil BDS 3 µm C18 150 mm x 2.1 mm i.d. columns. We achieved a good resolution of 15(R) 8-iso-PGF2
and 8-iso-PGF2
in standard solutions and urine samples using Hypersil BDS 3 µm C18 50 mm x 2.1 mm i.d. columns (data not shown), which confirms the finding of Li et al. (15). However, when urine samples were analyzed using Hypersil BDS 3 µm C8 columns, an additional peak was detected between 15(R) 8-iso-PGF2
and 8-iso-PGF2
and next to 8-iso-PGF2
(Fig. 4A). Our new finding is that an unknown F2-iP isomer was coeluted with 8-iso-PGF2
on C18 columns but was separated from 8-iso-PGF2
on C8 columns. Therefore, C8 columns were used in the novel HPLC method to give better specificity than C18 columns.
Because we found that buffer additives caused significant ion suppression of F2-iPs and the separation of F2-iP isomers was not affected by including acetic acid in the extracted urine samples (data not shown), an HPLC method was developed without the use of buffers to control pH. Alternatively, we used LC-MS-grade water for LC-MS/MS analysis. Buffer-free HPLC methods minimize the maintenance of the LC-MS/MS machine and save time required for the daily preparation of buffers. Only water, MeOH, and ACN were supplemented. The novel HPLC method is robust because slight variations in temperature, mobile phase composition, gradient steepness, pH, sample volume, and replacement of HPLC columns did not significantly affect the separation of the commercially available F2-iPs standards (Scheme 1) (data not shown). To demonstrate the robustness of the novel HPLC method, we also analyzed standard solutions and extracted urine samples in different HPLC machines with different system volumes (Waters Alliance 2796 and 2695 Separation Modules). Similar separation of F2-iPs was achieved by changing the prevolume parameter (300 and 600 µl for Waters Alliance 2796 and 2695 Separation Modules, respectively) and slightly adjusting the starting gradient organic content while keeping the gradient steepness constant (data not shown). This is the first report that a robust HPLC method for the separation of F2-iPs was efficiently developed using computer simulation software. This agrees with the finding of Snyder and Dolan (26) that the separation of stereoisomers as a function of temperature and gradient was predictable by a computer simulation using DryLab.
We found that the maintenance of HPLC columns and MS detectors could be minimized using the novel MD-SPE and HPLC methods. More than 400 injections of the extracted urine samples did not cause a significant increase in HPLC column backpressure (data not shown). Therefore, guard columns and precolumn filters were not used in the HPLC analysis. More than 40 injections of the extracted urine samples did not cause visible contamination of the sample cone in the MS detector. This not only saves time for MS maintenance but also reduces the variation in detection sensitivity. Also, our findings indicate that these novel MD-SPE and HPLC methods made possible the consistent recovery of IS of iPF2
-III and iPF2
-VI from urine samples and the nonsignificant matrix-related ion suppression for iPF2
-III and iPF2
-VI (Table 1).
One great advantage of the novel MD-SPE method is that the urine sample-processing time for LC-MS/MS analysis is short. We normally extract eight urine samples in 3 h. Because of the robustness of the MD-SPE method, it should not be difficult to increase sample throughput through automation. Also, the MD-SPE method can be used to process 10 ml of culture medium from cells, including endothelial cells and smooth muscle cells, or 1 or 2 ml urine samples from animals, including rabbits and rats, with only very slight modification [i.e., by replacing a 0.2 µm, 3 ml Captiva Filter cartridge with a 10 µm, 10 ml Captiva Filter cartridge (Varian, Inc.)] (data not shown).
However, the MD-SPE method will require major modifications if plasma samples are to be prepared for the analysis of F2-iPs using LC-MS/MS. Because F2-iPs are generated in situ esterified to phospholipids (1), alkaline hydrolysis of plasma samples is needed to cleave F2-iPs into free F2-iPs to measure total F2-iPs. Also, it may be worthwhile to add a free radical scavenger such as butylhydroxytoluene or an inhibitor of cyclooxygenase such as indomethacin (30) to EDTA plasma samples to prevent autoxidation during sample separation, storage, and processing. Iuliano et al. (31) added [2H8]arachidonic acid to plasma samples after collection to detect any artifactual formation of 8-iso-PGF2
-III. We found that retention of [3H]8-iso-PGF2
by Oasis HLB SPE cartridges was much lower in plasma samples that had been treated with 15% KOH in ethanol than in plasma samples that were not treated (data not shown). One milliliter of plasma sample overloaded the 30 mg/1 cc Oasis HLB cartridge. Therefore, Oasis HLB or Strata X SPE cartridges with a higher retention capacity (i.e., a greater sorbent amount such as 60 or 200 mg) are needed for plasma sample preparation. Therefore, the urine MD-SPE method has limitations in the analysis of F2-iPs in plasma samples.
In conclusion, we developed a novel sample clean-up method using MD-SPE and a novel HPLC method for the specific analysis of F2-iPs using LC-MS/MS. With this novel MD-SPE LC-MS/MS method, it should no longer be difficult to perform the routine analysis of F2-iPs in urine samples. Validation of the novel MD-SPE LC-MS/MS method for the routine analysis of iPF2
-III and iPF2
-VI in urine is now in progress in our laboratory.
| ACKNOWLEDGMENTS |
|---|
Manuscript received October 2, 2006 and in revised form November 2, 2006.
| REFERENCES |
|---|
|
|
|---|
| |||||||||||||||||||||||||||||||||||||||||||||||||||||